We have calibrated an LVDT using an optical and x-ray interferometer. We have calibrated optical interferometer using x-ray interferometer. The LVDT has calibrated by the optical interferometer. We made the monolithic x-ray interferometer with a double parallel spring structure for the translation of an analyzer lamella. One period of the xray interference fringe corresponds to the lattice parameter, 0.192 nm. The nonlinearity of optical interferometer has been calibrated by an x-ray interferometer. We have used a phase modulation optical interferometer. This calibration using the x-ray interferometer is directly traceable to primary standards. We have achieved the resolution of an xray interferometer and optical interferometer better than 0.01 nm. The optical phase stability of the interferometer is less than ± 150 pm. For the control of environmental temperature, we have used PID method. PID controller controlled the temperature inside chamber. Temperature drift was less than ± 3 mK (k = 2).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.