The present study was carried out to solve the problems of long washout time and non-linear calibration curves encountered in mercury analysis using inductively coupled plasma mass spectrometry. Comparisons of the washing efficiency for different reagents to eliminate the mercury memory effect were made. Of all the selected washing reagents, mercapto reagents such as 2-mercaptoethanol and L-cysteine could efficiently clean the instrument. Taking the toxicity and odor of mercaptoalkanol into account, L-cysteine was preferred as the most suitable washing reagent and was added into the standard and sample solutions. A good linear calibration curve was obtained with the correlation coefficient of 0.9999. The detetion limit for addition of L-cysteine was 0.024 mg l À1 . Addition of 0.18% L-cysteine into the sample solution also facilitates the washout of mercury even using deionized water. The recoveries of mercury in certified reference materials, i.e. human hair and dogfish muscle, were 97.2% and 98.3%, respectively, when using 0.18% L-cysteine in the sample solutions.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.