The use of cross-linkage in spinal fusion increases torsional stiffness in pedicle screw and hook constructs. This study 1) confirmed the effectiveness of cross-linkage in limiting torsional motion and showed the superiority of two cross-links to one cross-link in limiting torsional motion, 2) showed that increase of torsional stiffness of a cross-linked construct is proportional to the cross-sectional area of the cross-link, and 3) demonstrated that cross-links do not increases stiffness in the lateral flexion mode.
Caution is recommended when bending TI or CP rods, or placing a connector and removing it when using CP rods in situations in which fatigue failure is a concern.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.