Instrumented indentation test (IIT) is a depth-sensing hardness test allowing nano- to macro-mechanical characterisation of surface mechanical properties. Indenter tip geometry calibration allows nano-scale characterisation, overcoming the limits of conventional hardness tests. Calibration is critical to ensure IIT traceability and applicability for quality verification in manufacturing processes. The accuracy and precision of IIT are mainly affected by the indenter tip geometry calibration. State-of-the-art indenter tip geometry calibration reports either direct calibration by AFM, which is highly expensive and unpractical for industry, or indirect calibration methods, which are less accurate, precise and robust. This work proposes a practical, direct calibration method for IIT indenter tip geometry by optical surface topography measuring instruments. The methodology is complemented by uncertainty evaluation. The proposed approach is applied to Berkovich and Vickers indenters and its advantages are proven in terms of accuracy and precision of mechanical characterisation on metallic and ceramic material.
Graphical abstract
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.