Utrasonic backscatter is useful for characterizing tissues and several groups have reported methods for estimating backscattering properties. Previous interlaboratory comparisons have been made to test the ability to accurately estimate the backscatter coefficient (BSC) by different laboratories around the world. Results of these comparisons showed variability in BSC estimates but were acquired only for a relatively narrow frequency range, and, most importantly, lacked reference to any independent predictions from scattering theory. The goal of this study was to compare Faran-scattering-theory predictions with cooperatively-measured backscatter coefficients for low-attenuating and tissue-like attenuating phantoms containing glass sphere scatterers of different sizes for which BSCs can independently be predicted. Ultrasonic backscatter measurements were made for frequencies from 1 to 12 MHz. Backscatter coefficients were estimated using two different planar-reflector techniques at two laboratories for two groups of phantoms. Excellent agreement was observed between BSC estimates from both laboratories. In addition, good agreement with the predictions of Faran’s theory was obtained, with average fractional (bias) errors ranging from 8–14%. This interlaboratory comparison demonstrates the ability to accurately estimate parameters derived from the BSC, including an effective scatterer size and the acoustic concentration, both of which may prove useful for diagnostic applications of ultrasound tissue characterization.
Objective To translate quantitative ultrasound (QUS) from the laboratory into the clinic, it is necessary to demonstrate that the measurements are platform independent. Because the backscatter coefficient (BSC) is the fundamental estimate from which additional QUS estimates are calculated, agreement between BSC results using different systems must be demonstrated. This study was an intercomparison of BSCs from in vivo spontaneous rat mammary tumors acquired by different groups using 3 clinical array systems and a single-element laboratory scanner system. Methods Radio frequency data spanning the 1- to 14-MHz frequency range were acquired in 3 dimensions from all animals using each system. Each group processed their radio frequency data independently, and the resulting BSCs were compared. The rat tumors were diagnosed as either carcinoma or fibroadenoma. Results Carcinoma BSC results exhibited small variations between the multiple slices acquired with each transducer, with similar slopes of BSC versus frequency for all systems. Somewhat larger variations were observed in fibroadenomas, although BSC variations between slices of the same tumor were of comparable magnitude to variations between transducers and systems. The root mean squared (RMS) errors between different transducers and imaging platforms were highly variable. The lowest RMS errors were observed for the fibroadenomas between 4 and 5 MHz, with an average RMS error of 4 × 10−5 cm−1Sr−1 and an average BSC value of 7.1 × 10−4 cm−1Sr−1, or approximately 5% error. The highest errors were observed for the carcinoma between 7 and 8 MHz, with an RMS error of 1.1 × 10−1 cm−1Sr−1 and an average BSC value of 3.5 × 10−2 cm−1Sr−1, or approximately 300% error. Conclusions This technical advance shows the potential for QUS technology to function with different imaging platforms.
Utrasonic backscatter is useful for characterizing tissues and several groups have reported methods for estimating backscattering properties. Previous interlaboratory comparisons have been made to test the ability to accurately estimate the backscatter coefficient (BSC) by different laboratories around the world. Results of these comparisons showed variability in BSC estimates but were acquired only for a relatively narrow frequency range, and, most importantly, lacked reference to any independent predictions from scattering theory. The goal of this study was to compare Faranscattering-theory predictions with cooperatively-measured backscatter coefficients for lowattenuating and tissue-like attenuating phantoms containing glass sphere scatterers of different sizes for which BSCs can independently be predicted. Ultrasonic backscatter measurements were made for frequencies from 1 to 12 MHz. Backscatter coefficients were estimated using two different planar-reflector techniques at two laboratories for two groups of phantoms. Excellent agreement was observed between BSC estimates from both laboratories. In addition, good agreement with the predictions of Faran's theory was obtained, with average fractional (bias) errors ranging from 8-14%. This interlaboratory comparison demonstrates the ability to accurately estimate parameters derived from the BSC, including an effective scatterer size and the acoustic concentration, both of which may prove useful for diagnostic applications of ultrasound tissue characterization.
Applicability of ultrasound phantoms to biological tissue has been limited because most phantoms have generally used strong scatterers. The objective was to develop very weakly scattering phantoms, whose acoustic scattering properties are likely closer to those of tissues and then compare theoretical simulations and experimental backscatter coefficient (BSC) results. The phantoms consisted of agar spheres of various diameters (nominally between 90 and 212 microm), containing ultrafiltered milk, suspended in an agar background. BSC estimates were performed at two institutions over the frequency range 1-13 MHz, and compared to three models. Excellent agreement was shown between the two laboratory results as well as with the three models.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.