SummaryLaryngoscopy can induce stress responses that may be harmful in susceptible patients. We directly measured the force applied to the base of the tongue as a surrogate for the stress response. Force measurements were obtained using three FlexiForce Sensors Ò (Tekscan Inc, Boston, MA, USA) attached along the concave surface of each laryngoscope blade.Twenty-four 24 adult patients of ASA physical status 1-2 were studied. After induction of anaesthesia and neuromuscular blockade, laryngoscopy and tracheal intubation was performed using either a Macintosh or a GlideScope (93-207 [17-509]) Ns; p = 0.017) with the GlideScope. Our study shows that the peak lifting force on the base of the tongue during laryngoscopy is less with the GlideScope videolaryngoscope compared with the Macintosh laryngoscope.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.