Background Tension band wiring is the most common surgical procedure for fixation of fractures of the olecranon, but symptomatic hardware prominence and migration of K-wires can cause a high re-operation rate. The olecranon sled has been designed to minimize some of these problems.Material and methods Simulated olecranon fractures were created in 6 matched pairs of cadaver arms. Each pair was fixed with tension band wiring used on the one arm and the olecranon sled being used on the other. Mechanical testing was done with the humerus rigidly fixed in a vertical position while the forearm was held at 1 of 3 angles of elbow fixation, 45º, 90º and 135º, respectively. For each angle, the triceps and the brachialis muscles were sequentially loaded with 5 kg (50 N) for 20 cycles and the amount of fracture displacement measured.Results Loading of the brachialis muscle produced no increase in the fracture gap for either of the two fixation techniques. However, an increase in the fracture gap of up to 0.23 mm was found after cyclic loading of the triceps muscle for both techniques. The amount of increase was not significantly different between the two techniques.Interpretation The olecranon sled appears to provide as stable fixation as tension band wiring for olecranon fractures.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.