Time domain reflectometry (TDR) is used to measure moisture content and salinity of soils. Most TDR systems have a bandwidth of 2.5 GHz or less, limiting the precision of measurements using short probes. The primary objective of this research was to develop and test short probes for use with a high‐bandwidth (20‐GHz) TDR instrument. The secondary objective was to determine moisture content in highly conductive media by insulating the probes with Teflon heat‐shrinkable tubing. Laboratory‐packed test cells with probe lengths of 0.025, 0.05, and 0.075 m were used for the primary research objective, and 0.075‐m probes for the secondary objective. Linear relationships between the apparent refractive index and volumetric moisture content were developed for each probe length and type. The accuracy of the probes was ±0.0125, ±0.05, ±0.025, and ±0.035 m3/m3 for the probes of lengths 0.075, 0.075 (insulated), 0.05, and 0.025 m, respectively. Moisture measurements in highly conductive media were possible using insulated probes.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.