Background: Continuous, post-operative patient monitoring can provide healthcare workers with earlier warning signs of patient deterioration. The Masimo Pulse CO-Oximeter (Masimo Corporation, Irvine, CA) has acoustic monitoring technology which measures respiratory rate (RR) by analyzing acoustic signals generated across the upper airway while breathing. We continuously monitored for 19-24 hours (hrs) post-surgery RR and oxygen saturation (SpO2) in obese patients who underwent laparoscopic Roux-en-Y gastric bypass (RYGB) for bariatric surgery weight loss. The objective was to assess incidence of decreased RR (< 10 rpm), respiratory pauses and desaturation (< 90%) in a patient population that is at higher risk for hypoventilation, for an extended time after discharge from Post Anesthesia Care Unit (PACU). Methods: After Institutional Review Board approval, 20 patients aged 18-65 with a body mass index (BMI) between 40 kg/m 2 and 60 kg/m 2 who received primary laparoscopic RYGB surgery by the same surgeon were enrolled in this study. After surgery, post-operative pain management included IV patient-controlled anesthesia dilaudid as an adjunct to intraoperative lidocaine infusions (see reference 3). We recorded Visual Analogue Scale (VAS) scores every 3 hrs for 24 hrs. RR and SpO2 were continuously monitored from the time of PACU arrival and for the following 19-24 hrs. The SpO2 alarm was set at 90% and respiratory pause alarm was set at 30 seconds to ensure a rapid response. Floor nurses were alerted immediately when these values were reached. Collected data for SpO2 and RR were graphed versus time for the 24-hr period for each patient in order to evaluate the incidence of low values. The data were then graphed versus time to yield a representation of cumulative time spent at each SpO2 level or respiratory rate. The data were stratified into two ranges for SpO2 (SpO2 ≥ 90% and < 90%) and three ranges for RR (RR > 30, between 30 and 10, and < 10 breaths per minute). The percentage of time that the patient spent in each range was calculated. Median values were calculated for each category.Results: During the extended monitoring period, the median percentage of time spent with SpO2 below 90% was 5.4%. The median percentage of time that a subject had a RR of less than 10 respirations/min was 0%. There was no correlation between SpO2 and respiratory rate. Fourteen patients in our study population were on continuous positive airway pressure (CPAP) and reported VAS scores between 3 and 6. This may have prevented them from being deeply sedated. There were no apnea periods longer than 30 seconds. Conclusions: In this small group of obese patients, a low RR was not encountered during an extended post-operative period. These patients did not experience desaturation, apnea, or significant reductions of RR.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.