Background: To assess the relationship between mechanical wear and the failure of the internal lengthening mechanism in retrieved MAGnetic Expansion Control (MAGEC) growing rods. Methods: This study included 34 MAGEC rods retrieved from 20 patients. The state of the internal mechanism and mechanical wear were assessed in all the rods using plain radiographs and visual inspection. Metrology was then performed to assess the topography and mechanical wear of the telescopic bars, using a Talyrond 365 (Taylor Hobson, Leicester, UK) roundness measuring machine. Results: Plain radiographs showed evidence of a broken internal mechanism in 29% of retrieved rods. Single-side wear marks were found in 97% of retrieved rods. Material loss was found to significantly increase in rods with a damaged internal mechanism (p < 0.05) and rods with longer time in situ (r = 0.692, p < 0.05). Conclusions: We found an association between damage to the internal mechanism of the rods and (1) patterns of single-side longitudinal wear marks and (2) increased material loss. As the material loss was also found to increase over time of rod in situ, we emphasise the importance of early detection and revision of failed MAGEC rods in clinical practice.
Purpose To assess the reliability and validity of a handheld scanner (SpineScan3D) for trunk rotation measurement in adolescent idiopathic scoliosis (AIS) subjects, as compared with Scoliometer. Methods This was a cross-sectional study with AIS subjects recruited. Biplanar spine radiographs were performed using an EOS imaging system with coronal Cobb angle (CCA) determined. The angle of trunk rotation (ATR) was measured using Scoliometer. SpineScan3D was employed to assess the axial rotation of subjects’ back at forward bending, recorded as surface tilt angle (STA). Intra- and inter-examiner repeats were conducted to evaluate the reliability of SpineScan3D. Results 97 AIS patients were recruited. Intra- and inter-examiner reliability of STA measures were good to excellent in major thoracic and lumbar curves (p < 0.001). A strong correlation was found between STA and ATR measures in both curve types (p < 0.001) with a standard error of the ATR estimate of between 1 and 2 degrees from linear regression models (R squared: 0.8–0.9, p < 0.001). A similar correlation with CCA was found for STA and ATR measures (r: 0.5–0.6, p < 0.002), which also demonstrated a similar sensitivity (72%-74%) and specificity (62%-77%) for diagnosing moderate to severe curves. Conclusion SpineScan3D is a handheld surface scanner with a potential of wide applications in subjects with AIS. The current study indicated that SpineScan3D is reliable and valid for measuring trunk rotation in AIS subjects, comparable to Scoliometer. Further studies are planned to investigate its measurements in coronal and sagittal planes and the potential of this device as a screening and monitoring tool. Trial registration number (date of registration) HKUCTR-2288 (06 Dec 2017). Level of evidence Level III.
Background: To assess the relationship between mechanical wear and the failure of the internal lengthening mechanism in retrieved MAGnetic Expansion Control (MAGEC) growing rods. Methods: This study included 34 MAGEC rods retrieved from 20 patients. The state of the internal mechanism and mechanical wear were assessed in all the rods using plain radiographs and visual inspection. Metrology was then performed to assess the topography and mechanical wear of the telescopic bars, using a Talyrond 365 (Taylor Hobson, Leicester, UK) roundness measuring machine. Results: Plain radiographs showed evidence of a broken internal mechanism in 29% of retrieved rods. Single-side wear marks were found in 97% of retrieved rods. Material loss was found to significantly increase in rods with a damaged internal mechanism (p<0.05) and rods with longer time in situ (r=0.692, p<0.05). Conclusions: We found an association between damage to the internal mechanism of the rods and (1) patterns of single-side longitudinal wear marks and (2) increased material loss. As the material loss was also found to increase over time of rod in situ, we emphasise the importance of early detection and revision of failed MAGEC rods in clinical practice.
Purpose: To assess the relationship between mechanical wear and the failure of the internal lengthening mechanism in retrieved MAGnetic Expansion Control (MAGEC) growing rods. Methods: This study included 34 MAGEC rods retrieved from 20 patients. The state of the internal mechanism and mechanical wear were assessed in all the rods using plain radiographs and visual inspection. Metrology was then performed to assess the topography and mechanical wear of the telescopic bars, using a Talyrond 365 (Taylor Hobson, Leicester, UK) roundness measuring machine. Results: Plain radiographs showed evidence of a broken internal mechanism in 29% of retrieved rods. Single-side wear marks were found in 97% of retrieved rods. Material loss was found to significantly increase in rods with a damaged internal mechanism (p<0.05) and rods with longer time in situ (r=0.692, p<0.05). Conclusion: We found an association between damage to the internal mechanism of the rods and (1) patterns of single-side longitudinal wear marks and (2) increased material loss. As the material loss was also found to increase over time of rod in situ, we emphasise the importance of early detection and revision of failed MAGEC rods in clinical practice.
Background To assess the relationship between mechanical wear and the failure of the internal lengthening mechanism in retrieved MAGnetic Expansion Control (MAGEC) growing rods. Methods This study included 34 MAGEC rods retrieved from 20 patients. The state of the internal mechanism and mechanical wear were assessed in all the rods using plain radiographs and visual inspection. Metrology was then performed to assess the topography and mechanical wear of the telescopic bars, using a Talyrond 365 (Taylor Hobson, Leicester, UK) roundness measuring machine. Results Plain radiographs showed evidence of a broken internal mechanism in 29% of retrieved rods. Single-side wear marks were found in 97% of retrieved rods. Material loss was found to significantly increase in rods with a damaged internal mechanism (p<0.05) and rods with longer time in situ (r=0.692, p<0.05). Conclusion We found an association between damage to the internal mechanism of the rods and (1) patterns of single-side longitudinal wear marks and (2) increased material loss. As the material loss was also found to increase over time of rod in situ, we emphasise the importance of early detection and revision of failed MAGEC rods in clinical practice.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.