A simple, inexpensive and reliable manual spectrophotometric method for the determination of ozone in ambient air is evaluated. The air is passed through the blue solution of indigo disulphonate (IDS) in a special double sintered disc absorber. The resulting partly bleached solution is very stable. The discoloration is measured at 610 nm. The stoicheiometry of the reaction is independent of sample flow-rate and the amount of ozone absorbed. The collection efficiency is 100%, even at 2000 m l min-1. The decrease of absorbancecaused by ozone is 21 960 I mol-1 cm-1. Probably this value is largely independent of the brand and batch of IDS. Therefore, the method can be used for simple determinations without calibration when an inaccuracy up to about 20% is admissible. Otherwise, a calibration against a reference method is necessary.In an intercomparison study referenced against gas phase titration an apparent stoicheiometry factor of 1.13 was obtained. Probably this deviation from 1.00 was caused by non-reactive impurities in the IDS. Nitrogen dioxide gives a slightly positive interference (6% of its mass) but the other common air pollutants, in ambient concentration, do not interfere. The relative standard deviation for determinations of ozone concentrations between 50 and 1000 pg m-3 is less than 5%. The limit of detection is about 8 pg m-3.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.