This study describes a method of measuring the INR on native whole blood capillary samples using Innovin recombinant thromboplastin. Modification of the reagent was necessary to compensate for the nonoptimal level of calcium in the sample/reagent mixture. Ninety-five percent of results obtained by the capillary blood method were no more than 0.42 INR higher or 0.38 INR lower than the venous blood method. The effect of changes in haematocrit was minimal. Significant differences in results were found between the Innovin and Thrombotest capillary blood methods. Provided the reagent was properly stored, there was no reagent drift and satisfactory results were obtained on samples supplied by UKNEQAS (coagulation) from previous trials. The method described is a convenient, simple and accurate method of measuring the INR using native capillary whole blood and Innovin recombinant thromboplastin.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.