Cyclic physiological loading of osteoporotic olecranon fracture fixation resulted in sudden, catastrophic failure of the bone-implant interface rather than in gradual implant loosening. Recent plate innovations such as locking plates and different screw designs and positions appear to offer no advantages in stabilizing osteoporotic olecranon fractures. Surgeons may be reassured that the current olecranon plates will probably adequately stabilize osteoporotic fractures for early motion in the early postoperative period, but not for heavy activities such as those that involve over 4 kg of resistance.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.