A method for the determination of iodine value (IV) by Fourier transform-near infrared (FT-NIR) spectroscopy was developed and evaluated in an international collaborative study. The FT-NIR analyzer employed in this work uses disposable vials for sample handling and incorporates validation protocols designed to ensure that the calibration will give accurate results from analyzer to analyzer and stability over time without any further calibration development work. The global IV calibration was developed from over 1,200 animal, marine, and vegetable oils and fats, which were obtained on a number of different instruments worldwide. The Standard Error of Cross Validation measured from a range of 0-190 IV varied from ±0.2-1.4 IV (1 sigma). The repeatability for all models was on the order of 0.1 IV, which states that most of the error was inherited from the primary methods. Finally, an international interlaboratory study was carried out with 16 samples obtained from the AOCS Smalley Laboratory Proficiency Program and an oil processor. The average reproducibility error in any one lab was better than 0.15, and the average reproducibility between labs was better than 0.33. An uncertainty of 0.45 was calculated from the average FT-NIR values obtained from the collaborative study vs. the AOCS Certified Wijs method (Cd 1d-92).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.