Background—
The use of a stent to deliver a drug may reduce in-stent restenosis. Paclitaxel interrupts the smooth muscle cell cycle by stabilizing microtubules, thereby arresting mitosis.
Methods and Results—
On the basis of prior animal studies, the European evaLUation of the pacliTaxel Eluting Stent (ELUTES) pilot clinical trial (n=190) investigated the safety and efficacy of V-Flex Plus coronary stents (Cook Inc) coated with escalating doses of paclitaxel (0.2, 0.7, 1.4, and 2.7 μg/mm
2
stent surface area) applied directly to the abluminal surface of the stent in de novo lesions compared with bare stent alone. The primary efficacy end point was angiographic percent diameter stenosis at 6 months. At angiographic follow-up, percent diameter stenosis was 33.9±26.7% in controls (n=34) and 14.2±16.6% in the 2.7-μg/mm
2
group (n=31;
P
=0.006). Late loss decreased from 0.73±0.73 to 0.11±0.50 mm (
P
=0.002). Binary restenosis (≥50% at follow-up) decreased from 20.6% to 3.2% (
P
=0.056), with no significant benefit from intermediate paclitaxel doses. Freedom from major adverse cardiac events in the highest (effective) dose group was 92%, 89%, and 86% at 1, 6, and 12 months, respectively (
P
=NS versus control). No late stent thromboses were seen in any treated group despite clopidogrel treatment for 3 months only.
Conclusions—
Paclitaxel applied directly to the abluminal surface of a bare metal coronary stent, at a dose density of 2.7 μg/mm
2
, reduced angiographic indicators of in-stent restenosis without short- or medium-term side effects.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.