In 1980 an investigation was conducted due to an apparent cluster of fevers following transurethral resections of the prostate (TURP), where the presence of prostatitis or urinary tract infections (UTIs) could not account for most cases, and the usage of antimicrobials did not prevent them. When unused, prepackaged, sterilized samples from 13 lots of triple lumen catheters were analyzed, four were found to contain high levels of endotoxin (ET), with a range of 49,150-≥ 6.25 ng/catheter. Additionally, these lots were highly pyrogenic to rabbits. The maximum incidence of febrile patients (42%) correlated with usage of catheters from the lot where ET levels were the highest. The febrile rate for a second urologist, who used single lumen catheters, was less (12/33 v 0/12). Further consideration should be given to sterile operative equipment which may still be pyrogenic.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.