The improved Density Reference System, the reference densimeter, and the method of determining sample density are described. The uncertainty of the density reference system is ± 0.055%. The contribution from the estimated systematic error in density was ± 0.022%. The estimated uncertainty caused by random error is three times the standard deviation of 0.011% and is based on sixty-three measurements of the densities of saturated liquid methane. The total density uncertainty is taken to be the sum of the systematic 3 and random errors. This applies to the density range of 400 to 480 kg/m at pressures from 0.8 to 4 bar absolute and temperatures between 109 and 128 K. This 3 accuracy statement is expected to apply over ranges of at least 400 to 1000 kg/m in density, 77 to 300 K in temperature, and 0.8 to 7 bar in pressure though the accuracy over these ranges has not been verified. * Compared to the original, this improved density reference system has greater accuracy, speed, reliability, and flexibility.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.