When used by health professionals in a pharmacist-managed anticoagulation clinic, the CoaguChek S is reliable, valid and easy to use. However, its validity tends to decrease as the INR increases, possibly due to the low sensitivity of the thromboplastin. If the CoaguChek S INR is supratherapeutic, we would therefore recommend confirming the results with a standard laboratory measurement.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.