The positioning accuracy and precision of a head and neck immobilization device for radiation therapy of tumors in the canine skull was evaluated. Nineteen dogs with a spontaneous tumor of the head were enrolled including 12 with an intracranial mass and seven with an intranasal or maxillary tumor. Three hundred thirty-three pairs of orthogonal digital portal radiographs were analyzed to assess patient displacement in the cranial-caudal, lateral, and dorso-ventral directions. The mean systematic displacement was 0.8, 1, and 0.9 mm. The mean random displacement was 1.9, 1.6, and 1.5 mm. These values resulted in an overall displacement of 2.1 mm in the cranial-caudal direction, 1.8mm in the lateral direction, and 1.7 mm in the dorsal-ventral direction. The mean displacement value of the three dimensional (3D) vector was 2.4 mm with a standard deviation of 2.1. Ninety-five percent of all vectors were < 6.4 mm. This study quantifies the precision and accuracy of this particular positioning device. Knowing the limitations and setup variability of the system being used to set patients up for daily radiotherapy is paramount to planning and delivering appropriate radiation doses, especially as more complex treatment methods are used.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.