-In this paper, through a digital potentiometer and exponentially weighted moving average filter, pulse and PPG waveform measurable device was fabricated in radial artery. If this device is not proper about signal size in analog part, MCU can judge easily by adjusted amplification through digital potentiometer, using exponentially weighted moving average filter is able to filter out more clear value of ADC. I presumed pulse rate as value of measuring time between point of maximum contraction from sensing signal in radial artery of wrist. Therefore, this means can measure stable pulse rate and PPG waveform, finger as well as radial artery, whether signal size of each person is different finger as well as radial artery.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.