The aortic valve echogram was recorded in open chest dogs simultaneously with aortic flow, acceleration of flow, aortic and left ventricular pressures, and intracardiac phonocardiograms. Comparison of echographic with hemodynamic data showed the following: The aortic valve started its opening with the onset of flow, at the same point that left ventricular pressure exceeded aortic pressure. Complete valve opening preceded peak aortic flow by an average of 43 msec and the cusps started to move toward closure while flow was still accelerating. Final closure of the valve was achieved at the time of zero flow and preceded the aortic second sound by 4-10 msec. The echographically determined "valve orifice area" correlated well with aortic stroke volume (r = 0.94). The intensity of the aortic first sound was related to peak acceleration of aortic flow. The intensity of the second sound was not related to the amplitude of cusp motion but correlated well with the aortic pressure at the time of closure and with peak flow deceleration.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.