Independent verification of the production of element 114 in the reaction of 244-MeV ;{48}Ca with ;{242}Pu is presented. Two chains of time- and position-correlated decays have been assigned to ;{286}114 and ;{287}114. The observed decay modes, half-lives, and decay energies agree with published results. The measured cross sections at a center-of-target energy of 244 MeV for the ;{242}Pu(;{48}Ca,3-4n);{287,286}114 reactions were 1.4_{-1.2};{+3.2} pb each, which are lower than the reported values.
Purpose
To characterize the stacked and staggered dual‐layer multileaf collimator (MLC) on the HalcyonTM system.
Methods
The novel MLC assembly was reviewed and compared to the widely used MillenniumTM 120‐leaf MLC system. We investigated the MLC positioning stability over 70 days using Machine Performance Check (MPC) data. We evaluated the leaf transmission, penumbra, leaf end effect, and leaf edge effect. Leaf transmission through distal, proximal, and both MLC layers was measured with a Farmer chamber, by comparing an open and a closed field. Leaf penumbra was measured using film for three different MLC‐defined field sizes. The leaf end effect was measured with sweeping gap fields of varying gap sizes defined by the distal MLC. The leaf edge effect was evaluated using the Electronic Portal Imaging Device (EPID) for the different banks, gantry positions, and collimator angles. Point dose measurements for 10 test plans were compared to dose predictions of two dose calculation model versions.
Results
From MPC data, the largest measured MLC positioning accuracy deviation was within 0.1 mm. The proximal MLC exhibited greater deviations compared to the distal MLC. The distal‐and‐proximal‐combination had reduced inter‐leaf and intra‐leaf transmission compared to delivery with distal‐only. The measured leaf transmission was 0.41% for distal‐only, 0.40% for proximal‐only, and negligible for distal‐and‐proximal‐combination. The leaf end penumbra was wider compared to the leaf edge penumbra. The leaf end effect was measured to be −0.2 mm. The leaf edge effect showed minimal bank, gantry position, and collimator angle dependence. However, a systematic deviation between measurements and treatment planning system handling of the leaf edge effect was observed. The discrepancy between the measured and predicted dose in the 10 test plans improved with the latest version of the dose calculation algorithm.
Conclusion
The characteristics of the stacked and staggered dual‐layer MLC on the HalcyonTM system were presented.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.