Recent studies suggest ultra-high dose rate (FLASH) irradiation can spare normal tissues from radiotoxicity, while efficiently controlling the tumor, and this is known as the "FLASH effect". This study performed theoretical analyses about the impact of radiolytic oxygen depletion (ROD) on the cellular responses after FLASH irradiation.Methods: Monte Carlo simulation was used to model the ROD process, determine the DNA damage, and calculate the amount of oxygen depleted (LROD) during FLASH exposure. A mathematical model was applied to analyze oxygen tension (pO2) distribution in human tissues and the recovery of pO2 after FLASH irradiation. DNA damage and cell survival fractions (SFs) after FLASH irradiation were calculated. The impact of initial cellular pO2, FLASH pulse number, pulse interval, and radiation quality of the source particles on ROD and subsequent cellular responses were systematically evaluated.
Results:The simulated electron LROD range was 0.38-0.43 μM/Gy when pO2 ranged from 7.5-160 mmHg. The calculated DNA damage and SFs show that radioprotective effect is only evident in cells with a lower pO2. Different irradiation setups alter the cellular responses by modifying the pO2. Single pulse delivery or multi-pulse delivery with pulse intervals shorter than 10-50 ms resulted in fewer DNA damages and higher SFs. Source particles with a low radiation quality have a higher capacity to deplete oxygen, and thus, lead to a more conspicuous radioprotective effect.
Conclusions:A systematic analysis of the cellular response following FLASH irradiation was performed to provided suggestions for future FLASH applications. The FLASH radioprotective effect due to ROD may only be observed in cells with a low pO2. Single pulse delivery or multipulse delivery with short pulse intervals are suggested for FLASH irradiation to avoid oxygen tension recovery during pulse intervals. Source particles with low radiation quality are preferred for their conspicuous radioprotective effects.
Traditional data-driven respiratory gating method is capable of detecting breathing cycles directly from positron emission tomography (PET) data, but usually fails at low SNR, particularly at low dose PET/CT study. Time-of-flight (TOF) PET has the potential to improve the SNR. In order for TOF information to reduce the statistical noise and boost the performance of respiratory gating, we present a robust data-driven respiratory gating method using TOF information, which retrospectively derived the respiratory signal from the acquired TOF-PET data. The PET data was acquired in list mode format and analyzed in sinogram space. The method was demonstrated with patient datasets acquired on a TOF PET/CT system. Data-driven gating methods by center of mass (COM) and principle component analysis (PCA) algorithm were successfully performed on nonTOF PET and TOF PET dataset. To assess the accuracy of the data-driven respiratory signal, a hardware-based signal was acquired for comparison. The study showed that retrospectively respiratory gating using TOF sinograms has improved the SNR, and outperforms the non-TOF gating under both COM and PCA algorithms.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.