Cardiovascular diseases are the leading cause of death worldwide. Due to advancements facilitating the integration of electric and adhesive technologies, long-term patch electrocardiogram (ECG) monitors (PEMs) are currently used to conduct daily continuous cardiac function assessments. This paper presents an ECG encoding scheme for joint lossless data compression and heartbeat detection to minimize the circuit footprint size and power consumption of a PEM. The proposed encoding scheme supports two operation modes: fixed-block mode and dynamic-block mode. Both modes compress ECG data losslessly, but only dynamic-block mode supports the heartbeat detection feature. The whole encoding scheme was implemented on a C-platform and tested with ECG data from MIT/BIH arrhythmia databases. A compression ratio of 2.1 could be achieved with a normal heartbeat. Dynamic-block mode provides heartbeat detection accuracy at a rate higher than 98%. Fixed-block mode was also implemented on the field-programmable gate array, and could be used as a chip for using analog-to-digital convertor-ready signals as an operation clock.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.