It may be concluded from this study that although internal connections are clinically favored, this study did not show any advantage relative to screw loosening. However, screw design may be a significant factor in loosening of the joint.
Differences in screw design influence the maintenance of preload and distortion of the shank. The influence of the interface design, namely an internal hex of 1 mm versus an external hex did not influence the preload. Cantilevered prostheses can cause loss of torque and dimensional changes in abutment screws.
For both internal connection and external hexed systems, loss of screw tightness can be correlated with plastic deformation of the screw. This does not seem to be true, however, for a conical interface implant system.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.