Performing transcatheter tricuspid valve interventions requires a thorough knowledge of right-heart imaging. Integration of chamber views across the spectrum of imaging modalities (i.e., multislice computed tomography, fluoroscopy, and echocardiography) can facilitate transcatheter interventions on the right heart. Optimal fluoroscopic viewing angles for guiding interventional procedures can be obtained using pre-procedural multislice computed tomography scans. The present paper describes fluoroscopic viewing angles necessary to appreciate right-heart chamber anatomy and their relationship to echocardiography using multislice computed tomography.
Background
Recurrence of mitral regurgitation (MR) after surgical mitral valve repair (SMVR) varies and may require reoperation. Redo mitral valve surgery can be technically challenging and is associated with increased risk of mortality and morbidity. We aimed to assess the feasibility and safety of MitraClip as a treatment strategy after failed SMVR and identify procedure modifications to overcome technical challenges.
Methods and Results
This international multicenter observational retrospective study collected information for all patients from 16 high‐volume hospitals who were treated with MitraClip after failed SMVR from October 29, 2009, until August 1, 2017. Data were anonymously collected. Technical and device success were recorded per modified Mitral Valve Academic Research Consortium criteria. Overall, 104 consecutive patients were included. Median Society of Thoracic Surgeons score was 4.5% and median age was 73 years. At baseline, the majority of patients (82%) were in New York Heart Association class ≥III and MR was moderate or higher in 86% of patients. The cause of MR pre‐SMVR was degenerative in 50%, functional in 35%, mixed in 8%, and missing/unknown in 8% of patients. The median time between SMVR and MitraClip was 5.3 (1.9–9.7) years. Technical and device success were 90% and 89%, respectively. Additional/modified imaging was applied in 21% of cases. An MR reduction of ≥1 grade was achieved in 94% of patients and residual MR was moderate or less in 90% of patients. In‐hospital all‐cause mortality was 2%, and 86% of patients were in New York Heart Association class ≤II.
Conclusions
MitraClip is a safe and less invasive treatment option for patients with recurrent MR after failed SMVR. Additional/modified imaging may help overcome technical challenges during leaflet grasping.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.