Antibody-drug conjugates (ADCs) are a growing class of biotherapeutics in which a potent small molecule is linked to an antibody. ADCs are highly complex and structurally heterogeneous, typically containing numerous product-related species. One of the most impactful steps in ADC development is the identification of critical quality attributes to determine product characteristics that may affect safety and efficacy. However, due to the additional complexity of ADCs relative to the parent antibodies, establishing a solid understanding of the major quality attributes and determining their criticality are a major undertaking in ADC development. Here, we review the development challenges, especially for reliable detection of quality attributes, citing literature and new data from our laboratories, highlight recent improvements in major analytical techniques for ADC characterization and control, and discuss newer techniques, such as two-dimensional liquid chromatography, that have potential to be included in analytical control strategies.
Rationale:Multi-Attribute Methods (MAMs) are appealing due to their ability to provide data on multiple molecular attributes from a single assay. If fully realized, such tests could reduce the number of assays required to support a product control strategy while providing equivalent or greater product understanding relative to the conventional approach. In doing so, MAMs have the potential to decrease development and manufacturing costs by reducing the number of tests in a release panel.
Methods:In this work, we report a MAM which is based on subunit mass analysis.
Results:The MAM assay is shown to be suitable for use as a combined method for identity testing, glycan profiling, and protein ratio determination for coformulated monoclonal antibody (mAb) drugs. This is achieved by taking advantage of the high mass accuracy and relative quantification capabilities of intact mass analysis using quadrupole time-of-flight mass spectrometry (Q-TOF MS). Protein identification is achieved by comparing the measured masses of light chain (LC) and heavy chain (HC) mAbs against their theoretical values. Specificity is based on instrument mass accuracy. Glycan profiling and relative protein ratios are determined by the relative peak intensities of the protein HC glycoforms and LC glycoforms, respectively. Results for these relative quantifications agree well with those obtained by the conventional hydrophilic interaction liquid chromatography (HILIC) and reversed-phase LC methods.
Conclusions:The suitability of this MAM for use in a quality control setting is demonstrated through assessment specificity for mAb identity, and accuracy, precision, linearity and robustness for glycan profiling and ratio determination.Results from this study indicate that a MAM with subunit mass analysis has the potential to replace three conventional methods widely used for mAb release testing including identification assay, glycosylation profiling, and ratio determination for co-formulated mAbs.
A systematic approach was developed for identification of the color generating species in protein therapeutics and for estimation of the contribution of each color source.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.