To identify predictors of neonatal ECMO circuit health, a retrospective analysis of circuit functional pressure and flow parameters as well as infant clotting values were collected 48 h prior to and 24 h post circuit change. Circuit impairment was defined as need for partial or total circuit change. Statistical analysis used multivariate statistics and non-parametric Mann–Whitney U-test with possible non-normality of measurements. A total of 9764 ECMO circuit and clotting values in 21 circuits were analyzed. Circuit delta-P mean, and maximum values increased from 8.62 to 48.59 mmHg (p < 0.011) and 16.00 to 53.00 mmHg (p < 0.0128) respectively prior to need for circuit change. Maximum and mean Pump Flow Revolutions per minute (RPM) increased by 75% (p < 0.0043) and 81% (p < 0.0057), respectively. Mean plasma free hemoglobin (pfHb) increased from 26.45 to 76.00 mg/dl, (p < 0.0209). Sweep, venous pressure, and clotting parameters were unaffected. ECMO circuit delta-P, RPM, and pfHb were early predictors of circuit impairment.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.