IntroductionNew tests for improved diagnosis of diabetic peripheral neuropathy (DPN) are useful.Material and methodsWe evaluated the utility of automated nerve conduction study (NCS) of the sural nerve with a new portable device for the diagnosis of DPN in patients with type 2 diabetes mellitus (T2DM). This study included 114 T2DM patients (58 men) with mean age 64.60 ±8.61 years. Exclusion criteria were B12 depletion, alcohol abuse and other causes of peripheral neuropathy. The reference method was the Neuropathy Disability Score (NDS) with a threshold NDS ≥ 3. Sural nerve automated NCS was carried out with the portable NC-stat DPNCheck device. Sensory nerve conduction velocity and sensory nerve action potential amplitude were measured bilaterally. Automated NCS was considered abnormal when ≥ 1 of the two aforementioned neurophysiological parameters was abnormal in at least one leg.ResultsExamination with NC-stat DPNCheck exhibited 90.48% sensitivity, 86.11% specificity, 79.17% positive predictive value (PPV) and 93.94% negative predictive value (NPV). The positive likelihood ratio (LR+) was 6.51 and the negative likelihood ratio (LR–) was 0.11.ConclusionsSural nerve automated NCS with the NC-stat DPNCheck device exhibits high sensitivity and specificity for the diagnosis of DPN in T2DM.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.