We propose a new scanning protocol for generating 4D-CT image data sets influenced by respiratory motion. A cine scanning protocol is used during data acquisition, and two registration methods are used to sort images into temporal phases. A volume is imaged in multiple acquisitions of 1 or 2 cm length along the cranial-caudal direction. In each acquisition, the scans are continuously acquired for a time interval greater than or equal to the average respiratory cycle plus the duration of the data for an image reconstruction. The x ray is turned off during CT table translation and the acquisition is repeated until the prescribed volume is completely scanned. The scanning for 20 cm coverage takes about 1 min with an eight-slice CT or 2 mins with a four-slice CT. After data acquisition, the CT data are registered into respiratory phases based on either an internal anatomical match or an external respiratory signal. The internal approach registers the data according to correlation of anatomy in the CT images between two adjacent locations in consecutive respiratory cycles. We have demonstrated the technique with ROIs placed in the region of diaphragm. The external approach registers the image data according to an externally recorded respiratory signal generated by the Real-Time Position Management (RPM) Respiratory Gating System (Varian Medical Systems, Palo Alto, CA). Compared with previously reported prospective or retrospective imaging of the respiratory motion with a single-slice or multi-slice CT, the 4D-CT method proposed here provides (1) a shorter scan time of three to six times faster than the single-slice CT with prospective gating; (2) a shorter scan time of two to four times improvement over a previously reported multi-slice CT implementation, and (3) images over all phases of a breathing cycle. We have applied the scanning and registration methods on phantom, animal and patients, and initial results suggest the applicability of both the scanning and the registration methods.
Respiratory motion can introduce significant errors in radiotherapy. Conventional CT scans as commonly used for treatment planning can include severe motion artifacts that result from interplay effects between the advancing scan plane and object motion. To explicitly include organ/target motion in treatment planning and delivery, time-resolved CT data acquisition (4D Computed Tomography) is needed. 4DCT can be accomplished by oversampled CT data acquisition at each slice. During several CT tube rotations projection data are collected in axial cine mode for the duration of the patient's respiratory cycle (plus the time needed for a full CT gantry rotation). Multiple images are then reconstructed per slice that are evenly distributed over the acquisition time. Each of these images represents a different anatomical state during a respiratory cycle. After data acquisition at one couch position is completed, x rays are turned off and the couch advances to begin data acquisition again until full coverage of the scan length has been obtained. Concurrent to CT data acquisition the patient's abdominal surface motion is recorded in precise temporal correlation. To obtain CT volumes at different respiratory states, reconstructed images are sorted into different spatio-temporally coherent volumes based on respiratory phase as obtained from the patient's surface motion. During binning, phase tolerances are chosen to obtain complete volumetric information since images at different couch positions are reconstructed at different respiratory phases. We describe 4DCT image formation and associated experiments that characterize the properties of 4DCT. Residual motion artifacts remain due to partial projection effects. Temporal coherence within resorted 4DCT volumes is dominated by the number of reconstructed images per slice. The more images are reconstructed, the smaller phase tolerances can be for retrospective sorting. From phantom studies a precision of about 2.5 mm for quasiregular motion and typical respiratory periods could be concluded. A protocol for 4DCT scanning was evaluated and clinically implemented at the MGH. Patient data are presented to elucidate how additional patient specific parameters can impact 4DCT imaging.
Respiration-induced tumour motion can potentially compromise the use of intensity-modulated radiotherapy (IMRT) as a dose escalation tool for lung tumour treatment. We have experimentally investigated the intra-fractional organ motion effects in lung IMRT treatments delivered by multi-leaf collimator (MLC). An in-house made motor-driven platform, which moves sinusoidally with an amplitude of 1 cm and a period of 4 s, was used to mimic tumour motion. Tumour motion was simulated along cranial-caudal direction while MLC leaves moved across the patient from left to right, as in most clinical cases. The dose to a point near the centre of the tumour mass was measured according to geometric and dosimetric parameters from two five-field lung IMRT plans. For each field, measurement was done for two dose rates (300 and 500 MU min(-1)), three MLC delivery modes (sliding window, step-and-shoot with 10 and 20 intensity levels) and eight equally spaced starting phases of tumour motion. The dose to the measurement point delivered from all five fields was derived for both a single fraction and 30 fractions by randomly sampling from measured dose values of each field at different initial phases. It was found that the mean dose to a moving tumour differs slightly (<2-3%) from that to a static tumour. The variation in breathing phase at the start of dose delivery results in a maximum variation around the mean dose of greater than 30% for one field. The full width at half maximum for the probability distribution of the point dose is up to 8% for all five fields in a single fraction, but less than 1-2% after 30 fractions. In general, lower dose rate can reduce the motion-caused dose variation and therefore might be preferable for lung IMRT when no motion mitigation techniques are used. From the two IMRT cases we studied where tumour motion is perpendicular to MLC leaf motion, the dose variation was found to be insensitive to the MLC delivery mode.
Due to respiration, many tumours in the thorax and abdomen may move as much as 3 cm peak-to-peak during radiation treatment. To mitigate motion-induced irradiation of normal lung tissue, clinics have employed external markers to gate the treatment beam. This technique assumes that the correlation between the external surface and the internal tumour position remains constant inter-fractionally and intra-fractionally. In this work, a study has been performed to assess the validity of this correlation assumption for external surface based gated radiotherapy, by measuring the residual tumour motion within a gating window. Eight lung patients with implanted fiducial markers were studied at the NTT Hospital in Sapporo, Japan. Synchronized internal marker positions and external abdominal surface positions were measured during the entire course of treatment. Stereoscopic imaging was used to find the internal markers in four dimensions. The data were used retrospectively to assess conventional external surrogate respiratory-gated treatment. Both amplitude- and phase-based gating methods were investigated. For each method, three gating windows were investigated, each giving 40%, 30% and 20% duty cycle, respectively. The residual motion of the internal marker within these six gating windows was calculated. The beam-to-beam variation and day-to-day variation in the residual motion were calculated for both gating modalities. We found that the residual motion (95th percentile) was between 0.7 and 5.8 mm, 0.8 and 6.0 mm, and 0.9 and 6.2 mm for 20%, 30% and 40% duty cycle windows, respectively. Five of the eight patients showed less residual motion with amplitude-based gating than with phase-based gating. Large fluctuations (>300%) were seen in the residual motion between some beams. Overall, the mean beam-to-beam variation was 37% and 42% from the previous treatment beam for amplitude- and phase-based gating, respectively. The day-to-day variation was 29% and 34% from the previous day for amplitude- and phase-based gating, respectively. Although gating reduced the total tumour motion, the residual motion behaved unpredictably. Residual motion during treatment could exceed that which might have been considered in the treatment plan. Treatment margins that account for motion should be individualized and daily imaging should be performed to ensure that the residual motion is not exceeding the planned motion on a given day.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.