These results suggest that selection of device type and flow rate can influence vascular pulsatility and input impedance, which might affect clinical outcomes.
A need exists for a mock circulation that behaves in a physiologic manner for testing cardiac devices in normal and pathologic states. To address this need, an integrated mock cardiovascular system consisting of an atrium, ventricle, and systemic and coronary vasculature was developed specifically for testing ventricular assist devices (VADs). This test configuration enables atrial or ventricular apex inflow and aortic outflow cannulation connections. The objective of this study was to assess the ability of the mock ventricle to mimic the Frank-Starling response of normal, heart failure, and cardiac recovery conditions. The pressure-volume relationship of the mock ventricle was evaluated by varying ventricular volume over a wide range via atrial (preload) and aortic (afterload) occlusions. The input impedance of the mock vasculature was calculated using aortic pressure and flow measurements and also was used to estimate resistance, compliance, and inertial mechanical properties of the circulatory system. Results demonstrated that the mock ventricle pressure-volume loops and the end diastolic and end systolic pressure-volume relationships are representative of the Starling characteristics of the natural heart for each of the test conditions. The mock vasculature can be configured to mimic the input impedance and mechanical properties of native vasculature in the normal state. Although mock circulation testing systems cannot replace in vivo models, this configuration should be well suited for developing experimental protocols, testing device feedback control algorithms, investigating flow profiles, and training surgical staff on the operational procedures of cardiovascular devices.
This article presents an integrated model of the human circulatory system that incorporates circulatory support by a brushless DC axial flow ventricular assist device (VAD), and a feedback VAD controller designed to maintain physiologically sufficient perfusion. The developed integrated model combines a network type model of the circulatory system with a nonlinear dynamic model of the brushless DC pump We show that maintaining a reference differential pressure between the left ventricle and aorta leads to adequate perfusion for different pathologic cases, ranging from normal heart to left heart asystole, and widely varying physical activity scenarios from rest to exercise.
Flow and pressure measurements were performed in the ascending aortas of six pediatric patients ranging in age from 1 to 4 yr and in weight from 7.2 to 16.4 kg. From these measurements, input impedance was calculated. It was found that total vascular resistance decreased with increasing patient weight and was approximately one to three times higher than those of adults. Conductance per unit weight was relatively constant but was approximately three times higher than for adults. Strong inertial character was observed in the impedance of four of the six patients. Among a three-element and two four-element lumped-parameter models, the model with characteristic aortic resistor (R(c)) and inertance in series followed by parallel peripheral resistor (R(p)) and compliance fitted the data best. R(p) decreased with increasing patient weight and was one to three times higher than in adults, and R(c) decreased with increasing patient weight and was 2 to 15 times higher. The R(p)-to-R(c) ratio differed significantly between infants and children vs. adults. The results suggested that R(p) developed more rapidly with patient weight than did R(c). Compliance values increased with increasing patient weight and were 3 to 16 times lower than adult values.
An explanation of the mechanisms leading to the beneficial hemodynamic effects of the intra-aortic balloon pump (IABP) is lacking. We hypothesized that inflation and deflation of the balloon would generate a compression (BCW) and an expansion (BEW) wave, respectively, which, when analyzed with wave intensity analysis, could be used to explain the hemodynamic benefits of IABP support. Simultaneous ascending aortic pressure (Pao) and flow rate (Qao) were recorded in 25 patients during control conditions and with IABP support of 1:1 and 1:2. Diastolic aortic pressure augmentation (Paug) and end-diastolic aortic pressure (ED Pao) reduction were calculated from Pao. Energies of the BCW and BEW were obtained by integrating the wave intensity contour over time. Paug was 19.1 mmHg (SD 13.6) during 1:2 support. During 1:1 support significantly higher Paug of 21.1 mmHg (SD 13.4) was achieved (P < 0.001). ED Pao decreased from 50.9 mmHg (SD 15.1) to 43.9 mmHg (SD 15.7) (P < 0.0001) during 1:1 assistance and the decrease was not statistically different with 1:2. During 1:1 support the energy of BCW was correlated positively to Paug (r = 0.83, P < 0.0001) and energy of the BEW correlated negatively to ED Pao (r = 0.78, P < 0.005); these relationships were not statistically different during 1:2. In conclusion, the energies of the BCW and BEW are directly related to Paug and ED Pao, which are the conventional hemodynamic parameters indicating IABP benefits. These findings imply a cause and effect mechanism between the energies of BCW and BEW, and IABP hemodynamic effects.
Rotary blood pumps (RBPs) are currently being used as a bridge to transplantation as well as for myocardial recovery and destination therapy for patients with heart failure. Physiologic control systems for RBPs that can automatically and autonomously adjust the pump flow to match the physiologic requirement of the patient are needed to reduce human intervention and error, while improving the quality of life. Physiologic control systems for RBPs should ensure adequate perfusion while avoiding inflow occlusion via left ventricular (LV) suction for varying clinical and physical activity conditions. For RBPs used as left ventricular assist devices (LVADs), we hypothesize that maintaining a constant average pressure difference between the pulmonary vein and the aorta (deltaPa) would give rise to a physiologically adequate perfusion while avoiding LV suction. Using a mock circulatory system, we tested the performance of the control strategy of maintaining a constant average deltaPa and compared it with the results obtained when a constant average pump pressure head (deltaP) and constant rpm are maintained. The comparison was made for normal, failing, and asystolic left heart during rest and at light exercise. The deltaPa was maintained at 95 +/- 1 mm Hg for all the scenarios. The results indicate that the deltaPa control strategy maintained or restored the total flow rate to that of the physiologically normal heart during rest (3.8 L/m) and light exercise (5.4 L/m) conditions. The deltaPa approach adapted to changing exercise and clinical conditions better than the constant rpm and constant deltaP control strategies. The deltaPa control strategy requires the implantation of two pressure sensors, which may not be clinically feasible. Sensorless RBP control using the deltaPa algorithm, which can eliminate the failure prone pressure sensors, is being currently investigated.
This study investigated the hemodynamic and left ventricular (LV) pressure-volume loop responses to continuous versus pulsatile assist techniques at 50% and 100% bypass flow rates during simulated ventricular pathophysiologic states (normal, failing, recovery) with Starling response behavior in an adult mock circulation. The rationale for this approach was the desire to conduct a preliminary investigation in a well controlled environment that cannot be as easily produced in an animal model or clinical setting. Continuous and pulsatile flow ventricular assist devices (VADs) were connected to ventricular apical and aortic root return cannulae. The mock circulation was instrumented with a pressure-volume conductance catheter for simultaneous measurement of aortic root pressure and LV pressure and volume; a left atrial pressure catheter; a distal aortic pressure catheter; and aortic root, aortic distal, VAD output, and coronary flow probes. Filling pressures (mean left atrial and LV end diastolic) were reduced with each assist technique; continuous assist reduced filling pressures by 50% more than pulsatile. This reduction, however, was at the expense of a higher mean distal aortic pressure and lower diastolic to systolic coronary artery flow ratio. At full bypass flow (100%) for both assist devices, there was a pronounced effect on hemodynamic parameters, whereas the lesser bypass flow (50%) had only a slight influence. Hemodynamic responses to continuous and pulsatile assist during simulated heart failure differed from normal and recovery states. These findings suggest the potential for differences in endocardial perfusion between assist techniques that may warrant further investigation in an in vivo model, the need for controlling the amount of bypass flow, and the importance in considering the choice of in vivo model.
The acute hemodynamic efficacy of an implantable counter-pulsation device (CPD) was evaluated. The CPD is a valveless single port, 32-ml stroke volume blood chamber designed to be connected to the human axillary artery using a simple surface surgical procedure. Blood is drawn into the pump during systole and ejected during diastole. The acute hemodynamic effects of the 32-ml CPD were compared to a standard clinical 40-ml intra-aortic balloon pump (IABP) in calves (80 kg, n = 10). The calves were treated by a single oral dose of Monensin to produce a model of diminished cardiac function (DCF). The CPD and IABP produced similar increases in cardiac output (6% CPD vs. 5% IABP, p > 0.5) and reduction in left ventricular external work (14% CPD vs. 13% IABP, p > 0.5) compared to DCF (p < 0.05). However, the ratio of diastolic coronary artery flow to left ventricular external work increase from DCF baseline (p < 0.05) was greater with the CPD compared to the IABP (15% vs. 4%, p < 0.05). The CPD also produced a greater reduction in left ventricular myocardial oxygen consumption from DCF baseline (p < 0.05) compared to the IABP (13% vs. 9%, p < 0.05) despite each device providing similar improvements in cardiac output. There was no early indication of hemolysis, thrombus formation, or vascular injury. The CPD provides hemodynamic efficacy equivalent to an IABP and may become a therapeutic option for patients who may benefit from prolonged counterpulsation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.