Temperature rise, measured a short distance from a line heat source, can be used to determine the volumetric specific heat of soil and other materials. Volumetric specific heat is linearly related to the inverse of the temperature rise. The purpose of this note is to describe the construction and performance of a device for measuring specific heat using the line source method. The device was constructed from two hypodermic needles, 0.813 mm in diam. and 28 mm long, and spaced 6 mm apart. One needle contained a heater and the other a thermocouple. The temperature rise from heat pulses given to the heater were measured with the thermocouple. The coefficientof variation (CV) of specific heat on replicate samples was around 1%. Since water is the main variable component of the specific heat in nonswelling soil, changes in water content might be resolved to 0.01 or better in nonswelling soils.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.