Many different transducers are employed for recording respiratory sounds including accelerometers and microphones in couplers. However, there is no standard lung sound transducer or any device to compare transducers so that measurements from different laboratories can be determined to be of physiologic origin rather than technical artifacts of the transducers. To address this problem, we designed and constructed a prototype of a device that can be used to compare accelerometers, microphones enclosed in couplers, and stethoscopes. The prototype device consists of a rigid chamber containing a loudspeaker that opens to an antechamber covered by a viscoelastic material with mechanical properties similar to human skin and subcutaneous tissue. When driven by a white noise source, we found the sound output at the surface to be useful to comparatively evaluate sensors between 100 and 1200 Hz where lung sounds have most of their spectral energy. We compared the viscoelastic layer to similar thicknesses of fresh meat and fat and found them to produce similar acoustic spectra. This device allows air-coupled transducers, accelerometers, and stethoscopes used in respiratory sounds measurements to be compared under physical conditions similar to their intended use.
Sensors used for lung sound research are generally designed by the investigators or adapted from devices used in related fields. Their relative characteristics have never been defined. We employed an artificial chest wall with a viscoelastic surface and a white noise signal generator as a stable source of sound to compare the frequency response and pulse waveform reproduction of a selection of devices used for lung sound research. We used spectral estimation techniques to determine frequency response and cross-correlation of pulses to determine pulse shape fidelity. The sensors evaluated were the Siemens EMT 25 C accelerometer (Siemens); PPG 201 accelerometer (PPG); Sony ECM-T150 electret condenser microphone with air coupler (air coupler; with cylindrical air chambers of 5-, 10-, and 15-mm diameter and conical air chamber of 10-mm diameter); Littman classic stethoscope head (Littman) connected to an electret condenser microphone; and the Andries Tek (Andries) electronic stethoscope. We found that the size and shape of the air coupler chamber to have no important effect on the detected sound. The Siemens, air coupler, and Littman performed similarly with relatively flat frequency responses from 200 to 1,200 Hz. The PPG had the broadest frequency response, with useful sensitivity extending to 4,000 Hz. The Andries' frequency response was the poorest above 1,000 Hz. Accuracy in reproducing pulses roughly corresponded with the high-frequency sensitivity of the sensors. We conclude that there are important differences among commonly used lung sound sensors that have to be defined to allow the comparison of data from different laboratories.
Purpose: To determine whether nylon fibers improve the performance of platinum embolization coils in porcine arteries. Materials and Methods: Platinum 0.035" embolization coils, both with and without nylon fibers, were used to embolize a total of 24 hindlimb arteries in 6 swine: 12 with fibered coils and 12 with non-fibered coils. Apart from fibers, the coils were identical. Immediate and late results were studied, including number of coils needed to achieve vessel occlusion and durability of occlusion at 1 and 3 months. Arteriographic as well as histopathologic analysis were performed. Results: A mean of 3.2 (range, 2-4) non-fibered coils was required to achieve occlusion, whereas a mean of 1.3 (range, 1-2) fibered coils achieved occlusion in similarly sized arteries (2.3-3.2-mm diameter, P < .001). The mean percent cross-sectional area occupied by thrombus was greater in arteries with fibered coils than with non-fibered coils at 1 month (63% ± 6% and 48% ± 15%, respectively, P ¼ .03) but not at 3 months (61% ± 6% and 49% ± 15%, respectively, P ¼ .06). Some recanalization was observed at follow-up and did not differ between groups at 1 month (P ¼ .07) or 3 months (P ¼ .22). Conclusions: Nylon fibers allow significantly fewer embolization coils to achieve acute occlusion of arteries compared to bare metal coils. Both fibered and non-fibered coils showed recanalization at follow-up.
Several clinical and ambulatory settings necessitate respiratory monitoring without a mouthpiece or facemask. Several studies have demonstrated the utility of breathing sound measurements performed on the chest or neck to detect airflow. However, there are limitations to skin surface measurements, including susceptibility to external noise and transducer motion. Thus, this two-part study investigated a novel location for breathing sound measurements: the external ear. The first study investigated characteristics of sound transmission from the oropharynx to the external ear in 19 adults (nine males). Broadband noise was directed into the oropharynx through a tube and mouthpiece and measured indirectly via an accelerometer affixed to the cheek. Resultant transmission to the external ear was measured with a microphone inserted into an earplug that provided acoustic isolation from ambient noise. Near-unity coherence estimates (> 0.9) between the sounds recorded at the external ear and the oropharynx were observed up to approximately 800 Hz, indicating a low-frequency region of preferred transmission. In the second study, each of 20 subjects (nine males) breathed through a pneumotachograph at targeted shallow (3.0 mL/s/kg) and tidal (7.5 mL/s/kg) flows normalized to body mass, and the resulting sounds were recorded at the external ear. Recordings during breath hold measured background noise. Shallow and tidal expiratory flows, respectively, produced signal-plus-noise-to-noise [(S + N)/N] ratios of 6.7 +/- 4.1 dB and 14.0 +/- 5.3 dB (mean +/- standard deviation) across all subjects between 150 and 300 Hz. Concurrent inspiration demonstrated (S + N)/N ratios of 6.6 +/- 3.9 dB and 14.9 +/- 6.3 dB. Thus, the external ear shows promise as an anatomic site to detect and monitor breathing in a relatively noninvasive and unobtrusive manner.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.