Measured trajectories of markers on vessel wall and beads on a plumb are compared with calculated trajectories for transmission images obtained with Vista16™ optical cone beam scanner. Effective pixel size was 0.11 mm. An algorithm was developed to measure trajectories for the plumb line phantom and a semi-automatic version was required for the gel filled vessels due to overlap of markers and seam projections at maximum lateral positions. It was found that for 15 cm diameter vessels, the vessel edge was not visible due to divergence in cone beam geometry. Accounting for this effect was necessary in order to remove artefacts from calculated vessel trajectory. Displacement differences were found to be less than 0.1 mm for the plumb line, 0.3 mm for a beads and 1 mm for markers on gel filled vessel. Minimization of the vessel marker trajectory deviation will provide a more accurate approach to refractive index optimization. No optical aberrations were detected from the trajectory analysis and manufacturer scanner alignment and geometrical calibrations were independently verified.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.