This work aims to determine whether lead shielding can be used to decrease the radiation dose to the fetus during CT scans for the diagnosis of pulmonary embolism during early stage pregnancy. An anthropomorphic phantom was modified to contain a 15 cc ionization chamber at the site of the uterus to enable fetal dose to be measured. The effects of a range of scan parameters, positioning of lead and thicknesses of lead were investigated. Fetal dose was lower with lower values of kV(p) and mAs. An increasing thickness of lead decreased the radiation dose to the uterus, as did increasing the proportion of the patient covered by the lead shielding. Fetal dose increased exponentially as the edge of the scan volume moved closer to the point of measurement. In no experiment was the dose to the fetus increased by the presence of the lead. It was found that the fetal radiation dose from a CT scan following a pulmonary embolism protocol can be effectively reduced by the use of lead shielding.
Despite not being completely analogous investigations, computed tomography should be considered prior to angiography and not withheld on radiation dose concerns, given that it imparts lower and more consistent doses than conventional angiography.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.