This document is the report of a task group of the AAPM and has been prepared primarily to advise medical physicists involved in the external-beam radiation therapy of patients with thoracic, abdominal, and pelvic tumors affected by respiratory motion. This report describes the magnitude of respiratory motion, discusses radiotherapy specific problems caused by respiratory motion, explains techniques that explicitly manage respiratory motion during radiotherapy and gives recommendations in the application of these techniques for patient care, including quality assurance (QA) guidelines for these devices and their use with conformal and intensity modulated radiotherapy. The technologies covered by this report are motion-encompassing methods, respiratory gated techniques, breath-hold techniques, forced shallow-breathing methods, and respiration-synchronized techniques. The main outcome of this report is a clinical process guide for managing respiratory motion. Included in this guide is the recommendation that tumor motion should be measured (when possible) for each patient for whom respiratory motion is a concern. If target motion is greater than 5 mm, a method of respiratory motion management is available, and if the patient can tolerate the procedure, respiratory motion management technology is appropriate. Respiratory motion management is also appropriate when the procedure will increase normal tissue sparing. Respiratory motion management involves further resources, education and the development of and adherence to QA procedures.
Positron emission tomography (PET) has shown an increase in both sensitivity and specificity over computed tomography (CT) in lung cancer. However, motion artifacts in the 18F fluorodioxydoglucose (FDG) PET images caused by respiration persists to be an important factor in degrading PET image quality and quantification. Motion artifacts lead to two major effects: First, it affects the accuracy of quantitation, producing a reduction of the measured standard uptake value (SUV). Second, the apparent lesion volume is overestimated. Both impact upon the usage of PET images for radiation treatment planning. The first affects the visibility, or contrast, of the lesion. The second results in an increase in the planning target volume, and consequently a greater radiation dose to the normal tissues. One way to compensate for this effect is by applying a multiple-frame capture technique. The PET data are then acquired in synchronization with the respiratory motion. Reduction in smearing due to gating was investigated in both phantoms and patient studies. Phantom studies showed a dependence of the reduction in smearing on the lesion size, the motion amplitude, and the number of bins used for data acquisition. These studies also showed an improvement in the target-to-background ratio, and a more accurate measurement of the SUV. When applied to one patient, respiratory gating showed a 28% reduction in the total lesion volume, and a 56.5% increase in the SUV. This study was conducted as a proof of principle that a gating technique can effectively reduce motion artifacts in PET image acquisition.
We describe a method for generating CT images at multiple respiratory phases with a single spiral CT scan, referred to as respiratory-correlated spiral CT (RCCT). RCCT relies on a respiration wave form supplied by an external patient monitor. During acquisition this wave form is recorded along with the initiation time of the CT scan, so as to "time stamp" each reconstructed slice with the phase of the respiratory cycle. By selecting the appropriate slices, a full CT image set is generated at several phases, typically 7-11 per cycle. The CT parameters are chosen to optimize the temporal resolution while minimizing the spatial gap between slices at successive respiratory cycles. Using a pitch of 0.5, a gantry rotation period of 1.5 s, and a 180 degrees reconstruction algorithm results in approximately 5 mm slice spacing at a given phase for typical respiration periods, and a respiratory motion within each slice that is acceptably small, particularly near end expiration or end inspiration where gated radiotherapy is to occur. We have performed validation measurements on a phantom with a moving sphere designed to simulate respiration-induced tumor motion. RCCT scans of the phantom at respiratory periods of 4, 5, and 6 s show good agreement of the sphere's motion with that observed under fluoroscopic imaging. The positional deviations in the sphere's centroid between RCCT and fluoroscopy are 1.1+/-0.9 mm in the transaxial direction (average over all scans at all phases +/-1 s.d.) and 1.2+/-1.0 mm in the longitudinal direction. Reconstructed volumes match those expected on the basis of stationary-phantom scans to within 5% in all cases. The surface distortions of the reconstructed sphere, as quantified by deviations from a mathematical reference sphere, are similar to those from a stationary phantom scan and are correlated with the speed of the phantom. A RCCT scan of the phantom undergoing irregular motion, demonstrates that successful reconstruction can be achieved even with irregular respiration. Limitations from x-ray tube heating in our current CT unit restrict the length of the scan region to 9 cm for the RCCT settings used, though this will not be a limitation for a multislice scanner. RCCT offers an alternative to the current method of respiration-triggered axial scans. Multiple phases of respiration are imaged with RCCT in approximately the same scanning time required to image a single phase with a triggered axial scan. RCCT scans can be used in connection with respiratory-gated treatment to identify the patient-specific phase of minimum tumor motion, determine residual tumor motion within the gate interval, and compare treatment plans at different phases.
Abstract-Objective:Modeling of respiratory motion is important for a more accurate understanding and accounting of its effect on dose to cancers in the thorax and abdomen by radiotherapy. We have developed a model of respirationinduced organ motion in the thorax, without the commonly adopted assumption of repeatable breath cycles. Methods and Results
We have reported in our previous studies on the methodology, and feasibility of 4D-PET (Gated PET) acquisition, to reduce respiratory motion artifact in PET imaging of the thorax. In this study, we expand our investigation to address the problem of respiration motion in PET/CT imaging. The respiratory motion of four lung cancer patients were monitored by tracking external markers placed on the thorax. A 4D-CT acquisition was performed using a "step-and-shoot" technique, in which computed tomography (CT) projection data were acquired over a complete respiratory cycle at each couch position. The period of each CT acquisition segment was time stamped with an "x-ray ON" signal, which was recorded by the tracking system. 4D-CT data were then sorted into 10 groups, according to their corresponding phase of the breathing cycle. 4D-PET data were acquired in the gated mode, where each breathing cycle was divided into ten 0.5 s bins. For both CT and PET acquisitions, patients received audio prompting to regularize breathing. The 4D-CT and 4D-PET data were then correlated according to respiratory phase. The effect of 4D acquisition on improving the co-registration of PET and CT images, reducing motion smearing, and consequently increase the quantitation of the SUV, were investigated. Also, quantitation of the tumor motions in PET, and CT, were studied and compared. 4D-PET with matching phase 4D-CTAC showed an improved accuracy in PET-CT image co-registration of up to 41%, compared to measurements from 4D-PET with clinical-CTAC. Gating PET data in correlation with respiratory motion reduced motion-induced smearing, thereby decreasing the observed tumor volume, by as much as 43%. 4D-PET lesions volumes showed a maximum deviation of 19% between clinical CT and phase- matched 4D-CT attenuation corrected PET images. In CT, 4D acquisition resulted in increasing the tumor volume in two patients by up to 79%, and decreasing it in the other two by up to 35%. Consequently, these corrections have yielded an increase in the measured SUV by up to 16% over the clinical measured SUV, and 36% over SUV's measured in 4D-PET with clinical-CT Attenuation Correction (CTAC) SUV's. Quantitation of the maximum tumor motion amplitude, using 4D-PET and 4D-CT, showed up to 30% discrepancy between the two modalities. We have shown that 4D PET/CT is clinically a feasible method, to correct for respiratory motion artifacts in PET/CT imaging of the thorax. 4D PET/CT acquisition can reduce smearing, improve the accuracy in PET-CT co-registration, and increase the measured SUV. This should result in an improved tumor assessment for patients with lung malignancies.
We examine 2 strategies for reducing respiration-induced organ motion in radiation treatment: deep inspiration breath hold (DIBH) and respiratory gating. DIBH is a controlled breathing technique in which the patient performs a supervised breath hold during treatment. The technique offers 2 benefits: reduced respiratory motion from the breath hold and increased normal tissue sparing from the increased lung volume. In respiratory-gated treatment, a device external to the patient monitors breathing and allows delivery of radiation only during certain time intervals, synchronous with the patient's respiratory cycle. Gated treatment offers reduced respiratory motion with less patient effort than DIBH. We briefly survey the development of these 2 strategies, describe their clinical implementation for treatment of thoracic and liver tumors at the Memorial Sloan-Kettering Cancer Center, and discuss their advantages and limitations.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.