The measurement of the physical parameters of a high-energy proton beam, using a range of liquid and solid tissue substitutes, is described. The system, the detectors used and the experimental verification of the tissue equivalence of the new tissue substitutes is presented. The measurements with the scattered but uncollimated proton beam in muscle- and brain-equivalent liquids and in water are compared to similar data obtained from the scattered but collimated beam. The effect of lung, fat and bone on the dose distributions in composite phantoms is also investigated and the necessary corrections established. A simulated patient treatment indicated that the Bragg peak can be positioned with an error not exceeding +/- 0.5 mm.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.