This study investigated the effectiveness of a new method for measuring the actual focal spot position of a diagnostic x-ray tube using a high-precision antiscatter grid and a digital x-ray detector in which grid magnification, which is directly related to the focal spot position, was determined from the Fourier spectrum of the acquired x-ray grid’s image. A systematic experiment was performed to demonstrate the viability of the proposed measurement method. The hardware system used in the experiment consisted of an x-ray tube run at 50 kVp and 1 mA, a flat-panel detector with a pixel size of 49.5 µm, and a high-precision carbon-interspaced grid with a strip density of 200 lines/inch. The results indicated that the focal spot of the x-ray tube (Jupiter 5000, Oxford Instruments) used in the experiment was located approximately 31.10 mm inside from the exit flange, well agreed with the nominal value of 31.05 mm, which demonstrates the viability of the proposed measurement method. Thus, the proposed method can be utilized for system’s performance optimization in many x-ray imaging applications.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.