Infrequent control, aging of components, may compromise the accuracy of ICU ventilators. In order to assess the reliability of ventilators during their clinical use, we bench tested a group of 20 CPU1 ventilators (Ohmeda) sampled at random in several ICU units. We found major leaks in 5 ventilators, attributable to the disposable tubings used in these systems. Mean error in expired tidal volume and corresponding standard deviation (precision) were greater than 100 ml in two. Positive end expiratory pressure measurement comprised a mean error higher than 2 cm H2O in 40% of the ventilators tested. The valve opening pressure threshold was correlated to the inspiratory flow (r = 0.81) contrary to the valve opening delay (average 138 +/- 40 ms). These two parameters did not correlate with the age of the ventilator. Our study addresses the need for periodic control of ventilator performance in order to minimize the risks of errors and malfunctions.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations鈥揷itations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.