Purposes
To primarily evaluate MRI-induced effects for Ultra 3D cochlear implantation in human cadavers in terms of artifact generation and MR image quality.
Methods
Three human cadaveric heads were submitted to imaging after unilateral and bilateral cochlear implantation. The 1.5 T MR examination protocol was chosen in accordance with our institutional protocol for the assessment of brain pathology. The maximal signal void size was measured according to each sequence and plane. Two experienced neuro-radiologists and one experienced otoneurosurgeon independently evaluated the MR image quality findings. A 4-point scale was used to describe the diagnostic usefulness of 14 brain structures.
Results
Shape and size of the artifacts were found to be highly related to MRI sequences and acquisition planes. MRI sequences and processing algorithms affected the ability to assess anatomical visibility. Image quality appeared either high or assessable for diagnostic purposes in 9 out of 14 of the ipsilateral structures, in at least one plane. Anatomical structures contralateral to the cochlear implant were highly visible in all conditions. Artifact intrusion clearly improved after application of metal artifact-reduction techniques. In the case of bilateral cochlear implant, a mutual interaction between the two implant magnets produced an additional artifact.
Conclusions
We performed the first cadaver study aimed at systematically evaluating the MRI-induced artifacts produced by a cochlear implant with a novel four bar magnet system. Specific brain structures can be assessable for diagnostic purposes under 1.5 T MRI, with the cochlear implant magnet in place.
Purposes
To investigate the effects for Ultra 3D cochlear implant (CI) positioning on MR imaging quality, looking at a comprehensive description of intracranial structures in cases of unilateral and bilateral CI placement.
Methods
Four CI angular positions (90°, 120°, 135° and 160°) at 9 cm distance from the outer-ear canal were explored. The 1.5 T MRI assessment included our institutional protocol for the investigation of brain pathologies without gadolinium application. Three investigators (two experienced neuroradiologists and one experienced otoneurosurgeon) independently evaluated the MR findings. A 4-point scale was adopted to describe 14 intracranial structures and to determine which CI positioning allowed the best image quality score and how bilateral CI placement modified MRI scan visibility.
Results
A high positive correlation was found between the three blinded observers. Structures situated contralateral from the CI showed high-quality values in all four placements. Structures situated ipsilaterally provided results suitable for diagnostic purposes for at least one position. At 90°, artifacts mainly involved brain structures located cranially and anteriorly (e.g., temporal lobe); on the contrary, at 160°, artifacts mostly influenced the posterior fossa structures (e.g., occipital lobe). For the bilateral CI condition, MR imaging examination revealed additional artifacts involving all structures located close to either CI, where there was a signal void/distortion area.
Conclusions
Suitable unilateral CI positioning can allow the visualization of intracranial structures with sufficient visibility for diagnostic purposes. Bilateral CI positioning significantly deteriorates the anatomical visibility. CI positioning might play a crucial role for patients who need post-operative MRI surveillance.
In our preliminary report, translabyrinthine vestibular nerve section and simultaneous cochlear implant showed encouraging results in order to definitively control both vestibular and cochlear symptoms during the same therapeutic procedure.
The Journal of International Advanced Otology (J Int Adv Otol) is an international, peer reviewed, open access publication that is fully sponsored and owned by the European Academy of Otology and Neurotology and the Politzer Society. The journal is published triannually in April, August, and December and its publication language is English.The scope of the Journal is limited with otology, neurotology, audiology (excluding linguistics) and skull base medicine.The Journal of International Advanced Otology aims to publish manuscripts at the highest clinical and scientific level. J Int Adv Otol publishes original articles in the form of clinical and basic research, review articles, short reports and a limited number of case reports. Controversial patient discussions, communications on emerging technology, and historical issues will also be considered for publication.Target audience of J Int Adv Otol includes physicians and academics who work in the fields of otology, neurotology, audiology and skull base medicine.
SUMMARY
Introduction
The retrosigmoidal (RS) placement of the Bonebridge system (BB) has been advocated for cases of unfavourable anatomical or clinical conditions which contraindicate transmastoid-presigmoidal positioning. However, these disadvantageous conditions, combined with the considerable dimensions of the implant, may represent a challenge, especially for surgeons with no skull base experience. Moreover, the literature reports only limited experience concerning RS implantation of the BB system.
Methods
A multicentre, retrospective study was conducted to analyse the surgical and functional outcomes of a wide population of patients undergoing RS placement of the BB system by means of a surgical technique specifically developed to overcome the intraoperative issues related to this surgery. Twenty patients with conductive or mixed hearing loss and single sided deafness were submitted to RS implantation of the BB system.
Results
Audiological assessment concerning the measurement of the functional and effective gain by pure-tone audiometry (28 dB HL and -12.25 dB HL, respectively) and speech audiometry (24.7 dB HL and -21 dB HL, respectively) was conducted. A high overall subjective improvement of quality of life was recorded with the Glasgow Benefit Inventory questionnaire. No major complications, such as device extrusions or other conditions requiring revision surgery, were reported during the follow-up period (median: 42 months).
Conclusions
In our study, which has one of the largest cohort of patients reported in the literature, RS placement of the BB system was safe and effective. Our functional results showed comparable hearing outcomes with presigmoidal placement. The effective gain, rarely investigated in this field, may be the object of further research to improve our understanding of bone conduction mechanisms exploited by bone conduction hearing implants.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.