A home-use device that allows rapid and quantitative sperm quality analysis is desirable but not yet fully realized. To aid this effort, this article presents a microfluidic device capable of quantifying sperm quality in terms of two critical fertility-related parameters-motile sperm concentration and motility. The microdevice produces flow field for sperms to swim against, and sperms that overcame the flow within a specified time are propelled along in a separate channel and counted via resistive pulse technique. Data are compared to two control methods clinically utilized for sperm quality exam-hemocytometer and the sperm quality analyzer. Results reveal the numbers of pulses generated by passage of sperms correlates strongly with the two control methods: pulse number from 0 to 335 corresponds to progressively motile sperm concentrations from 0 to 19 9 10 6 / ml (hemocytometer) and Sperm Motility Index from 0 to 204 (sperm quality analyzer). The microdevice should be applicable to facilitate self-assessment of sperm quality at home.
likely an under diagnosed condition. While ejaculatory dysfunction is known to accompany alpha blocker use, the association with ED suggests a population who can be queried for symptoms and offered treatment options given that many men are bothered by DE.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.