This paper reports a novel method for the realization of small differential pressures in the range of 1 Pa and 1 kPa by using a precision wind tunnel with Laser Doppler Anemometer (LDA) and a Prandtl tube. The LDA system measures the wind speed of the precisely controlled wind tunnel and provides traceability to SI. A Prandtl tube immersed into the wind tunnel, relates the well-controlled airflow speed to the pressure difference of total pressure and static pressure at the point of measurement. Experiments have been performed in the range between 0.5 and 60 m/s air velocity corresponding to about 150 mPa and 2.2 kPa. The measurement uncertainty of the generated differential pressures between 1 Pa and 1 kPa are calculated to be around 0.8% (k = 2).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.