To improve the voice quality of laryngectomised patients, a voice-producing element has been developed. Prior to in vivo testing we constructed and validated an in-vitro test set-up, consisting of a physical model of the subglottal tract and three physical models of the vocal tract, for the vowels /a/, /i/ and /u/ to evaluate the voice-producing element under physiologic acoustic conditions. To meet acoustic conditions described in the literature, we determined the appropriate dimensions of these physical models, using a numerical model of the pressure perturbation in rigid tubes. The numerical model showed that an acoustic equivalent of the subglottal tract could be obtained with a three-tube system and an end impedance. Vocal tract models could be constructed using two- and four-resonator tubes. The physical models were built and successfully validated according to the human acoustic properties. The developed in-vitro set-up can now be applied to test voice-producing elements or vocal fold models under physiologic acoustic conditions.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.