A new device measuring water removal during standard dialysis is evaluated. The filtrate, collected from a small hemofilter inserted into a normal Cuprophan hollow-fiber dialyzer, was used to evaluate the total water removed from the patient. The device was tested in 46 patients undergoing regular dialysis treatment; the body weight loss ranged from 300 to 5,600 ml for a total of 71 dialysis sessions. Results confirmed the reliability of the device, as the mean prediction error was 5.4%. No influence of the dialyzer blood rest volume on the prediction error was observed. The authors propose this system as an alternative to bed or armchair scales and emphasize its usefulness for experimental purposes.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.