The electrocardiogram (ECG) interpretation in patients with implantable cardioverter defibrillator (ICD) is often a puzzling problem. The difficulty of the device function evaluation further increases in the presence of unfamiliar timing cycles and additional functions. Authors present a special function of a Biotronik ICD devices called the thoracic impedance monitoring, and demonstrate its behavior in a patient with sinus rhythm, ventricular ectopic beats, and ventricular tachycardia episode. This report shows unexceptional occurrence of tricky ECG finding in patient with Biotronik ICD.
The electrocardiogram (ECG) interpretation in patients with implantable cardioverter defibrillator (ICD) is often a puzzling problem. The difficulty of the device function evaluation further increases in the presence of unfamiliar timing cycles and additional functions.
We present an interesting ECG with a special function of a Biotronik ICD devices called the thoracic impedance monitoring, and demonstrate its behavior in a patient with atrial fibrillation, pacing beats, ventricular ectopic beats, and couple of ventricular beats. This report shows unexceptional occurrence of tricky ECG finding in patient with Biotronik ICDs.
The electrocardiogram (ECG) interpretation in patients with implantable pacemaker is often a perplexing problem. The difficulty in the device evaluation increases in the presence of novel timing cycles and additional functions. Authors describe a special function frequently encountered in Medtronic dual-chamber pacemakers and implantable cardioverter-defibrillator devices called managed ventricular pacing (MVP) and demonstrate its performance in the patient with undersensing episodes in ventricular channel. Intermittent ventricular undersensing in the device with MVP feature turned on caused repetitive mode switches between AAI and DDD mode. This report shows unexceptional occurrence of tricky ECG findings in patient with Medtronic dual-chamber device.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.