A coronary CTA protocol using 100 kVp tube voltage maintained image quality, but reduced radiation exposure by 31% as compared with the standard 120 kVp protocol. Thus, 100 kVp scan protocols should be considered for nonobese patients to keep radiation exposure as low as reasonably achievable. (Prospective Randomized Trial on Radiation Dose Estimates of Cardiac CT Angiography in Patients Scanned With a 100 kVp Protocol [PROTECTION II]; NCT00611780).
Using a scanner with a collimation of 16 x 0.625 mm, our study confirms the potential role of MSCT in the detection of significant coronary stenosis with a sensitivity of 89% and a very high specificity (98%). Exclusion criteria and less than full evaluability of the coronary arteries must still be considered limitations of the method.
This large international radiation dose survey demonstrates considerable reduction of radiation exposure in coronary CTA during the last decade. However, the large inter-site variability in radiation exposure underlines the need for further site-specific training and adaptation of contemporary cardiac scan protocols.
Palmaz-Schatz stent implantation after successful balloon PTCA of chronic total occlusions improves the midterm angiographic and clinical outcome and could be the preferred treatment option in selected patients with occluded vessels.
Objective
To determine whether coronary computed tomography angiography (CTA) should be performed in patients with any clinical probability of coronary artery disease (CAD), and whether the diagnostic performance differs between subgroups of patients.
Design
Prospectively designed meta-analysis of individual patient data from prospective diagnostic accuracy studies.
Data sources
Medline, Embase, and Web of Science for published studies. Unpublished studies were identified via direct contact with participating investigators.
Eligibility criteria for selecting studies
Prospective diagnostic accuracy studies that compared coronary CTA with coronary angiography as the reference standard, using at least a 50% diameter reduction as a cutoff value for obstructive CAD. All patients needed to have a clinical indication for coronary angiography due to suspected CAD, and both tests had to be performed in all patients. Results had to be provided using 2×2 or 3×2 cross tabulations for the comparison of CTA with coronary angiography. Primary outcomes were the positive and negative predictive values of CTA as a function of clinical pretest probability of obstructive CAD, analysed by a generalised linear mixed model; calculations were performed including and excluding non-diagnostic CTA results. The no-treat/treat threshold model was used to determine the range of appropriate pretest probabilities for CTA. The threshold model was based on obtained post-test probabilities of less than 15% in case of negative CTA and above 50% in case of positive CTA. Sex, angina pectoris type, age, and number of computed tomography detector rows were used as clinical variables to analyse the diagnostic performance in relevant subgroups.
Results
Individual patient data from 5332 patients from 65 prospective diagnostic accuracy studies were retrieved. For a pretest probability range of 7-67%, the treat threshold of more than 50% and the no-treat threshold of less than 15% post-test probability were obtained using CTA. At a pretest probability of 7%, the positive predictive value of CTA was 50.9% (95% confidence interval 43.3% to 57.7%) and the negative predictive value of CTA was 97.8% (96.4% to 98.7%); corresponding values at a pretest probability of 67% were 82.7% (78.3% to 86.2%) and 85.0% (80.2% to 88.9%), respectively. The overall sensitivity of CTA was 95.2% (92.6% to 96.9%) and the specificity was 79.2% (74.9% to 82.9%). CTA using more than 64 detector rows was associated with a higher empirical sensitivity than CTA using up to 64 rows (93.4%
v
86.5%, P=0.002) and specificity (84.4%
v
72.6%, P<0.001). The area under the receiver-operating-characteristic curve for CTA was 0.897 (0.889 to 0.906), and the diagnostic performance of CTA was slightly lower in women than in with men (area under the curve 0.874 (0.858 to 0.890)
v
0.907 (0.897 to 0.916), P<0.001). The diagnostic performance of CTA was slightly lower in patients older than 75 (0.864 (0.834 to 0.894), P=0.018
v
all other age groups) and was not significantly influenced by angina pectoris type (typical angina 0.895 (0.873 to 0.917), atypical angina 0.898 (0.884 to 0.913), non-anginal chest pain 0.884 (0.870 to 0.899), other chest discomfort 0.915 (0.897 to 0.934)).
Conclusions
In a no-treat/treat threshold model, the diagnosis of obstructive CAD using coronary CTA in patients with stable chest pain was most accurate when the clinical pretest probability was between 7% and 67%. Performance of CTA was not influenced by the angina pectoris type and was slightly higher in men and lower in older patients.
Systematic review registration
PROSPERO CRD42012002780.
In patients with stable and low heart rates, the prospectively ECG-triggered axial scan protocol maintained image quality but reduced radiation exposure by 69% compared with helical scanning. Axial computed tomography data acquisition should be strongly recommended in suitable patients to avoid unnecessarily high radiation exposure.
Background-Computed tomography has been shown to be useful in the evaluation of aortocoronary bypass grafts (CABG). This is the first prospective study to evaluate the accuracy of a new-generation scanner in the detection of patency and significant stenoses (Ͼ50% decrease in diameter) of venous and arterial grafts in patients with previous CABG. Methods and Results-In 96 patients (80 males, mean age 62 years) with previous CABG, a multislice computed tomography (MSCT) scan was performed (collimation 16ϫ0.625 mm). Patients with atrial fibrillation, renal failure, severe respiratory disease, severe heart failure, heart rate Ͼ70 bpm despite therapy, or unstable angina were excluded. A total of 285 conduits implanted on the native coronary arteries at the time of CABG were evaluated. MSCT data were analyzed by 2 independent radiologists and compared with the results of conventional angiography. Three patients were excluded from analysis. All conduits were judged evaluable in 84 patients. Among these patients, MSCT correctly diagnosed 54 occluded grafts and 4 significant stenoses on the body of the grafts. Of the 17 significant anastomotic lesions, MSCT correctly diagnosed 15. For these 84 patients, diagnostic accuracy was 99%, sensitivity was 97%, and specificity was 100%. When all 93 patients were considered, the sensitivity of MSCT in diagnosing significant stenoses was 96%. Conclusions-MSCT with the new-generation scanner allows for accurate assessment of venous and arterial conduits in patients with previous CABG with a high degree of sensitivity and specificity. Exclusion criteria and radiation exposure remain limitations of the method.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.