IRE can generate thermal energy, and even thermal injury, based on tissue type, probe exposure lengths, pulse lengths and proximity to metal. Awareness of probe placement regarding proximity to critical structures as well as probe exposure length and pulse length are necessary to ensure safety and prevent thermal injury. A probe exposure of 2·5 cm or less for liver IRE, and 1·5 cm or less for pancreas, with maximum pulse length of 90 µs will result in safe and non-thermal energy delivery with spacing of 1·5-2·3 cm between probe pairs.
SVV can be used safely as an alternative to CVP monitoring during hepatic resection with equivalent outcomes in terms of blood loss and parenchymal transection time. Using SVV as a predictor of fluid status could prove to be advantageous by avoiding the need for CVC insertion and therefor eliminating the risk of CVC related complications in patients undergoing hepatic resection.
In pancreatic adenocarcinoma patients with LN disease, survival after PD is comparable regardless of HALN status. Therefore, HALN-positive disease should not preclude the performance of PD.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.