Purpose
To compare subbasal nerve densities estimated from images recorded by the Tandem Scanning and the ConfoScan 4 confocal microscopes.
Methods
Confocal microscopy was used to estimate subbasal nerve density in 62 corneas of 40 subjects (18 corneas of 18 normal subjects and 44 corneas of 22 patients between 1 and 12 months after LASIK) At each examination, corneas were scanned first by using Tandem Scanning and then by using a ConfoScan 4 confocal microscope. Subbasal nerves from 2 to 4 scans per cornea were traced by using a semi-automated nerve analysis program. Nerve density was expressed as total nerve length divided by the sample area (µm/mm2). Differences in nerve density between instruments were examined by using paired tests.
Results
In normal corneas, subbasal nerve density was 10,658 ± 5,581 µm/mm2 (mean ± SD) with the ConfoScan 4 and 5,534 ± 1,850 µm/mm2 with the Tandem Scanning microscope (P<0.0001). One to 12 months after LASIK, mean subbasal nerve density was 2,477 ± 3,514 µm/mm2 estimated with the ConfoScan 4 and 844 ± 983 µm/mm2 estimated with the Tandem Scanning (P=0.0003). Estimates of nerve density were correlated between instruments (r=0.71, P<0.0001), although the mean difference between instruments was 2,308 ± 3,885 µm/mm2 (P<0.0001).
Conclusions
Mean subbasal nerve density estimated with the ConfoScan 4 was 2 to 3 times higher than density estimated with the Tandem Scanning confocal microscope. These differences must be considered when comparing subbasal nerve densities between studies that use different confocal microscopes.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.