Background:
There are rare reports of broken surgical blades occurring during lumbar discectomy, and even fewer that discuss their retrieval.
Case Description:
While a 54-year-old male was undergoing a lumbar discectomy, the knife blade was broken. As it was difficult to retrieve the fragment through the original incision, the patient was closed, and a postoperative angio-computerized tomography (CT) was obtained. When the CT angiogram (CTA) documented the retained fragment had become lodged near the iliac vein within the psoas muscle, a second operation for blade retrieval, consisting of a paravertebral, lateral transpsoas approach, was successfully performed.
Conclusion:
In some cases, it is difficult to retrieve a broken scalpel blade during the index surgery. When this occurs, we would recommend closing the patient, and obtaining a CTA to better document the location of the retained foreign body. Based upon these findings, a safer second stage procedure may be performed (e.g., as in this case using a paravertebral lateral transpsoas approach) to avoid undue sequelae/morbidity.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.