In this paper we present a technique to measure the radius of curvature of a test sphere based on the relation between acquired images of a circular cosine fringe pattern and size of virtual image formed on the calibration test surface. Radius of curvature is calculated with the exact equation proposed, using the parameters of the optical setup. Fringe pattern evaluation was performed by locating extrema indices. The mathematical formulation as well as the experimental setup and results are presented. After applying a linear fit algorithm to the data as a method of compensation, obtained results show an error within the tolerance established by the ISO 10343 specifications.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.